The PT-104 is a four channel temperature measuring data logger. It offers the ultimate in resolution (0.001 °C) and accuracy (0.01 °C) when used with PT100 sensors. It can also be used to measure resistance and voltage.

pt104

  • Measures temperature, resistance and voltage
  • High resolution and accuracy
  • No power supply required
  • For use with PT100 and PT1000 sensors

A wide range of PT-100 sensors are available for use with the PT-104.

Accuracy and Resolution

Although accurate temperature sensors are widely available, it has been difficult to take advantage of them due to errors caused by the measuring device. The PT104 however, is inherently accurate due to its novel design. Rather than relying on voltage references (which tend to be temperature sensitive) it uses 'reference' resistors which are extremely stable (low temperature co-efficient and drift). The exact value of each resistor is stored in an EEPROM to provide the ultimate in accuracy (yearly re-calibration is recommended). To achieve the 0.001 °C resolution a state of the art ADC is used that can resolve to better than 1 part in 16 million.

Temperature

The PT104 measures temperature using platinum resistance thermometers (PRTs). Both common industry standards (PT100 and PT1000) are supported. The unit is compatible with 2, 3 and 4 wire sensors (4 wire PT100 sensors are recommended for accurate measurements). More information on sensors can be found below.

Resistance

When measuring resistance, the unit uses a four-wire circuit to give the greatest possible accuracy.

Two resistance ranges are available (0 to 375 Ohms and 0 to 10,000 Ohms). The unit is calibrated for 0 to 375 Ohms so this range should be used for accurate measurements.

Voltage

For voltage measurements, each input connector can be treated as a differential input with ground, or two single ended inputs. Both inputs must be zero Volts or above, though it does not matter which input has the higher voltage.

Two voltage ranges are available (0 to 115 mV and 0 to 2500 mV). For accurate measurements use the 0 to 2500 mV range.

Software

The PT-104 is supplied with PicoLog data logging software for advanced data logging. PicoLog will automatically detect which sensor is connected and will display readings in the correct units. For users who wish to write their own custom software, drivers and examples are supplied.

Remote Data Collection

The PT-104 is normally connected directly to a PC, but it is also possible to communicate with the PT-104 using a modem (radio or telephone) so that you can collect data from a remote site.

 

PT-104 Specifications Temperature Resistance Voltage
Sensor PT100*, PT1000 N/A N/A
Range -200 to 800 °C 0 to 375 Ω*
0 to 10 kΩ
0 to 115 mV
0 to 2.5 V*
Linearity 10 ppm 10 ppm 10 ppm
Accuracy @25 °C 0.01 °C* 20 ppm* 0.2%*
Temperature coefficient 3 ppm/°C 3 ppm/°C 100 ppm/°C
RMS Noise (using filter) 0.01 °C 10 ppm 10 ppm
Resolution 0.001 °C 1 µΩ 0.156 µV
Conversion time per channel 720 mS ** 720 mS ** 180 mS
Number of inputs 4
Connectors 4-pin miniDIN
Input impedence >>1 MΩ
Overvoltage protection ±100 V
Output RS232, D9 female
Environmental 20 to 30 °C for stated accuracy, 0 to 70 °C overall, 20 to 90% RH
Software PicoLog data acquisition software for Windows (98SE/ME/XP/2000)
Drivers & examples C, pascal, Delphi, Visual Basic, HP VEE and LabVIEW. A macro is also provided to collect data directly into an Excel spreadsheet
Quoted accuracy is for options marked *
** For four wire temperature and resistance measurement