| Specifications |
HI
991001 |
HI
991002 |
| Ranges: |
pH |
-2.00 to 16.00
|
-2.00 to 16.00 |
| |
ORP |
- |
±1999 mV |
| |
Temperature |
-5.0 to 105.0°C |
-5.0 to 105.0°C |
| Resolution: |
pH |
0.01 |
0.01 |
| |
ORP |
- |
1 mV |
| |
Temperature |
0.1°C |
0.1°C |
| Accuracy |
pH |
±0.02 |
±0.02 |
| (@20°C/68°F): |
ORP |
- |
±2 mV |
| |
Temperature |
±0.5 up to 60°C,
±1°C up to 105°C
|
±0.5 up to 60°C,
±1°C up to 105°C
|
| Typical EMC |
pH |
±0.02 |
±0.02 |
| Deviation: |
ORP |
- |
±2 mV |
| |
Temperature |
±0.2°C |
±0.2°C |
| pH Calibration |
Automatic 1 or 2
points with 2 set of standardized buffers |
| |
(4.01, 7.01,
10.01 or 4.01, 6.86, 9.18) |
| Temp. Compensation: |
Automatic from -5
to 105°C (for pH only) |
| Electrode: |
HI 1296D combination
amplified pH/temperature
|
HI 1297D combination
amplified pH/ORP/temperature
|
| |
with DIN connector and 1
m
(3.3’) cable (included)
|
with DIN connector and 1
m
(3.3’) cable (included)
|
| Environment: |
0 to 50°C (32 to
122°F); RH 100% |
| Battery Type: |
3 x 1.5V AA |
| Battery Life: |
approximately
1500 hours of continuous use.
Auto-off after 8 minutes
|
| Dimensions: |
150 x 80 x 36 mm
(5.9 x 3.1 x 1.4”) |
| Weight: |
245 g (8.6 oz.) |